Thermowell Thermocouple

Thermowell Thermocouple

Our Thermowell Thermocouple is used to measure temperature in pressurized pipelines, steam systems and industrial process equipment. It is commonly installed in thermal power plants, boiler systems, chemical reactors and high-pressure process lines. We also offer thermowell RTD assemblies for applications that require resistance-based temperature measurement.
Send Inquiry
Description

Our Thermowell Thermocouple is used to measure temperature in pressurized pipelines, steam systems and industrial process equipment. It is commonly installed in thermal power plants, boiler systems, chemical reactors and high-pressure process lines. We also offer thermowell RTD assemblies for applications that require resistance-based temperature measurement.

 

The assembly uses a protective thermowell fitted to the process equipment, with the sensing element installed inside it. The thermowell stays in place when the sensor is withdrawn for inspection or replacement. This allows maintenance teams to service the measuring element without removing the protective housing from the pipeline. We supply different thermowell designs for general industrial service and power station applications.

 

Core Advantages

 

Replaceable sensing element: The sensor can be withdrawn from the installed thermowell, allowing maintenance without removing the pipeline connection.

Forged protective structure: The power station series uses a deep blind-hole thermowell structure, avoiding welded joints in the main protective body.

High-pressure suitability: The design supports pressures up to 30 MPa, making it suitable for demanding steam measurement points.

Material flexibility: 321 stainless steel is available for standard conditions, while GH3030 can be selected for higher-temperature service.

Different response configurations: The 624 insulated-junction and 625 grounded-junction designs provide alternative response characteristics.

0dffd0872bf8a01aaf8d676de3cb2854
060dfdeb3d2ad13013b228807042d7e5

Applications

 

  • Main steam pipelines in thermal power plants.
  • High-pressure boiler drums and steam systems.
  • Chemical reactors operating under elevated pressure.
  • Industrial auxiliary steam pipelines.
  • Power generation equipment requiring sensor replacement without pipeline removal.

OEM/ODM Customization

 

We produce Thermowell Thermocouple assemblies according to pipeline drawings and operating parameters. Customization includes thermowell length, insertion depth, pressure rating, connection dimensions, material grade, thermocouple type, RTD element and terminal configuration. We also offer 01T, 624 and 625 series options. For export projects, material certificates, English documentation and applicable product conformity documents can be prepared according to the confirmed order.

7893923ca85329763ea1a0d666abd744
2026062010453918058

Quality Control

 

Our manufacturing inspection covers protective tube material, structural dimensions, pressure-related requirements and sensing element performance. Material identification and finished-product checks are carried out before shipment. Our group has CNAS-accredited testing laboratories and a dedicated technical team supporting customized instrumentation. Third-party inspection can be coordinated according to the agreed inspection plan.

 

Product Specifications

 

Parameter

General-Purpose Thermowell Type (01T Series)

Power Station Type (624/625 Series)

Application

Small and medium-sized boilers and high-pressure auxiliary equipment pipelines

Main steam pipelines in large thermal power plants

Structural Design

Detachable assembly with spring-loaded clamping structure

Deep blind-hole forged construction with spring-loaded clamping structure

Common Thermocouple Types

K-type thermocouple / Pt100 RTD

K-type thermocouple / Pt100 RTD

Temperature Range

Thermocouple: 0°C to 800°C; RTD: -200°C to 500°C

Thermocouple: 0°C to 800°C; RTD: -200°C to 500°C

Maximum Pressure Rating

≤30 MPa

≤30 MPa

Applicable Flow Velocity

≤100 m/s

≤100 m/s

Standard Material

1Cr18Ni9Ti (321 stainless steel)

1Cr18Ni9Ti (321 stainless steel)

High-Temperature Material

GH3030 nickel-based alloy

GH3030 nickel-based alloy

Measuring Junction Design

General-purpose contact type

624: insulated junction; 625: grounded junction (faster response)

Explosion-Proof Version

WRN-014T (Ex d IIC T6)

WRN-6244 (Ex d IIC T6)

Representative Models

WRN-01T, WZP-01T, WRN2-01T

WRN-624, WRN-625, WZP-624

Applicable Standards

Domestic power industry standards, EBASCO standards

Domestic power industry standards, EBASCO standards

Warranty Period

18 months

18 months

 

FAQ

 

Q: Can the sensor be replaced while the pipeline remains in operation?

A: The removable sensing element allows replacement without removing the thermowell. Whether replacement can be performed while the process remains pressurized depends on the specific design, isolation arrangement and operating procedure.

Q: What is the difference between the 624 and 625 models?

A: The 624 version uses an insulated measuring junction, while the 625 version uses a grounded junction with a faster response characteristic.

Q: Can you manufacture thermowells for 30 MPa applications?

A: Yes. Our thermowell series supports a maximum specified pressure of 30 MPa, subject to the confirmed design and operating conditions.

 

Hot Tags: thermowell thermocouple, China thermowell thermocouple manufacturers, suppliers, factory, Bi Metal Temperature Gauge, digital panel meter, Explosion proof Bimetallic Thermometer, temperature measurement controller, temperature sensing instrument, Wear Resistant Thermocouple and Abrasion Resistant RTD

Send Inquiry